MODIFICATION

66 -- Air Monitoring Benches

Notice Date
7/5/2012
 
Notice Type
Modification/Amendment
 
NAICS
541380 — Testing Laboratories
 
Contracting Office
CPODUS Environmental Protection Agency26 West Martin Luther King DriveMail Code: NWDCincinnatiOH45268USA
 
ZIP Code
00000
 
Solicitation Number
SOL-CI-12-00053
 
Response Due
7/10/2012
 
Archive Date
8/9/2012
 
Point of Contact
William Yates
 
Small Business Set-Aside
N/A
 
Description
Questions and Answers for SOL-C-12-00053 1) Question: What are the stack parameters to be measured estimated concentrations? Answer: See the tables below for the low and high ranges and minimum detectable limits for each gas for Benches #1 & #2. Bench #1 (Stack) GasLow RangeHigh RangeMDL Multi-Gas Analyzer O5%25%0.01% CO100 ppm2500 ppm1 ppm CO25%25%500 ppm NO100 ppm2000 ppm1 ppm NO250 ppm500 ppm0.2 ppm SO2100 ppm10000 ppm0.2 ppm NO,NO2,NOx0.05 ppm100 ppm0.4 ppb Bench #2 (Ambient) GasLow RangeHigh RangeMDL Ozone Generatorozone output: 4.75 gr/hr O3 Analyzer0.05 ppm200 ppm O3 CalibratorSupports instruments on Bench 2 Zero Air System10 lpm20 lpm SO2 Analyzer10 ppb1000 ppb0.05 ppb NO,NO2,Nox5 ppb1000 ppb50 ppt CO Analyzer1 ppm10000 ppm0.04 ppm Multi-gas Dynamic Dilution CalibratorSupports instruments on Bench 2 2) Question: What is the temperature and pressure at the sample point? Answer: Ambient temperature and pressure. Sample lines may require pulling samples from chambers running at -20? H2O (negative 20?) vacuum level. 3) Question: Will we need to provide a sample probe? Answer: No If so please provide the stack information: Stack Outside Diameter ? Not applicable Stack Wall Thickness ?Not applicable Flange Size ? Not applicable 4) Question: What is your instrument air pressure at site? Answer: Instrument air pressure is available up to 90 psi. 5) Question: Do you prefer a Dilution or Full Extractive System? Answer: Full extractive system. 6) Question: Will we need to provide a sample probe if so please provide sample point information: Answer: No 7) Question: Please provide me detail on what capabilities you require from the Human Monitor Interface (HMI) terminal? Answer: We need an Ethernet net (TCP/IP) connection to interface with our own control and data system and either a 0-1 v, 0 ? 5 v, or 0-10 v DC analog output for each instrument. 8) Question: For the list of analyzers on both Bench #1 and #2, please provide the range requirements for each component Answer: See previous table for question 1. 9) Question: From the description within the bid document, the analyzers for both Bench #1 and #2 shall be installed in dual bay racks. Please confirm the desired arrangement for each system. Answer: Bench 1 instruments will be mounted in dual side by side racks. Instruments will be slide mounted. Rack will include casters for portability. Bench 2 instruments will be mounted in one vertical rack. Instruments will be slide mounted. Rack will include casters for portability. 10) Question: From the bid document, it appears that you do not require any plumbing, wiring of the system or data acquisition system. Please confirm that the limit of delivery of both systems will be the inlet sample ports of each analyzer. Answer: The limit of the delivery of both systems will be the inlet sample port, the Ethernet connection and the DC volt analog output. 11) Question: Please confirm the requirement for "installation, acceptance testing, and training" as noted in the last paragraph of the bid request document. Answer: Installation: No on-site installation is required. Acceptance testing: The instruments must operate according to the manufacturer?s specifications. Training: One day of training is requested; approximately one half day for each bench. Training should include setup, routine operation, calibration, data interface operation, and routine maintenance procedures. 12) Question: Will there be a more detailed specification coming? Answer: No the specifications are as stated. Additional Information: Additionally, we require submission for each instrument type of 1 each instrument manual, 1 each instrument plumbing (internal and external connections) schematic, 1 each instrument external wiring connection schematic and plug/terminal configuration. These documents should be submitted within 2 weeks of receipt of order. The due date for all quotes have been extended to 4:30 PM, EDT, July 24, 2012. Submit quote via email to: yates.william@epa.gov and to davies.david@epa.gov with subject line: "Air Quality Monitoring System, SOL-C-12-00053. Telephone requests for information may be honored.
 
Web Link
FBO.gov Permalink
(https://www.fbo.gov/spg/EPA/OAM/OH/SOL-CI-12-00053/listing.html)
 
Record
SN02795057-W 20120707/120705235447-8c3381d14054759f02a5bff43029b7df (fbodaily.com)
 
Source
FedBizOpps Link to This Notice
(may not be valid after Archive Date)

FSG Index  |  This Issue's Index  |  Today's FBO Daily Index Page |
ECGrid: EDI VAN Interconnect ECGridOS: EDI Web Services Interconnect API Government Data Publications CBDDisk Subscribers
 Privacy Policy  Jenny in Wanderland!  © 1994-2024, Loren Data Corp.