SOLICITATION NOTICE

66 -- Two auto-titration systems, including warranty service coverage and installation service

Notice Date
6/5/2020 6:50:27 AM
 
Notice Type
Combined Synopsis/Solicitation
 
NAICS
334516 — Analytical Laboratory Instrument Manufacturing
 
Contracting Office
USDA, FSIS, OAS PCMB BELTSVILLE MD 20705 USA
 
ZIP Code
20705
 
Solicitation Number
123A9420Q0016
 
Response Due
6/11/2020 8:00:00 AM
 
Archive Date
06/26/2020
 
Point of Contact
Thomas Nasvytis
 
E-Mail Address
thomas.nasvytis@usda.gov
(thomas.nasvytis@usda.gov)
 
Description
Please see the questions on the last two pages of the Statement of Work.� Questions about this solicitation are due at 11:00am (EST) on June 8, 2020. The laboratory requires two auto-titration systems, each with a pH titrator and an auto-sampler to accommodate tubes that are ? 50 mL in size. The aminoglycosides method requires an adjustment of each sample extract to a pH of 7.50 � 0.25, so the pH meter has to have an accuracy of at least 10E-2. In combination, the two auto-titration systems need to be able to analyze up to 48 samples total. The contractor shall supply the lab with two fully integrated auto-titration systems that are capable of titrating up to 24 samples per system, within the same run, with minimal interferences. The system shall have a ? 24 sample auto-sampler rack, capable of holding �?50 �mL tubes. Digital electrodes will produce data that will be read automatically into the integrated instrument software. Fixed ground-joint diaphragm electrodes will prevent contamination. The auto-titration systems will include rods and stirring propellers to streamline mixing. The auto-titration systems shall be compact and each fit on an approximately 30� wide existing lab bench/table top. The auto-titration systems should include all necessary supplies and accessories for operation and should be able to connect to a sing computer controller via USB drives. The instruments shall be installed and all factory published specifications for sensitivity demonstrated upon installation. The contractor shall provide warranty service coverage for the auto-titration instruments for at least one year after installation. The warranty service shall include telephone support for hardware and software and supply for the consumable parts during repairs and maintenance.� Maintenance/repair services shall be provided during normal working hours, 8:00 am to 4:30 pm, local time, Monday through Friday, excluding Federal Holidays.�
 
Web Link
SAM.gov Permalink
(https://beta.sam.gov/opp/6c49c6e4a2de4afcb5b5278f8a1e8c0d/view)
 
Place of Performance
Address: Albany, CA 94710, USA
Zip Code: 94710
Country: USA
 
Record
SN05682295-F 20200607/200605230147 (samdaily.us)
 
Source
SAM.gov Link to This Notice
(may not be valid after Archive Date)

FSG Index  |  This Issue's Index  |  Today's SAM Daily Index Page |
ECGrid: EDI VAN Interconnect ECGridOS: EDI Web Services Interconnect API Government Data Publications CBDDisk Subscribers
 Privacy Policy  Jenny in Wanderland!  © 1994-2024, Loren Data Corp.