COMMERCE BUSINESS DAILY ISSUE OF OCTOBER 28,1998 PSA#2210National Institutes of Health, Office of Procurement Management, 6011
Executive Blvd., Rockville, Maryland 20892-7260 66 -- ACOMA MOBILE HIGH FREQUENCY C-ARM HFC-9 WITH DIGIMAX SYSTEM SOL
263-99-B(GQ)-0005 DUE 111098 POC Dorothy Nickens,Contract Specialist,
(301) 496-0301, Fax 3 (301) 402-3407 NIH intends to negotiate and
award, on a sole-source basis, a contract to provide Acoma Mobile High
Frequency C-Arm HFC-9 with Digimax Digital System with Reliant Medical
Systems, Inc., Hanover, MD. This action is taken under the authority of
41 U.S.C. 253(c ) (1) (FAR 6.301-1(b). The salient
characteristics/performance features for Acoma Mobile High Frequency
C-Arm HFC-9 with Digimax Digital System shall have a high frequency
generator with at least 7.5 KW output, time selections shall range from
.032 seconds to 5.0 seconds, shall have anatomically programmed
radiography, shall feature a high resolution digital CCD camera with
last image hold and noise reduction, the C-arm assembly shall offer
optimum positioning with a variety of angular projections. (rotation
+/- 180 degree, orbital rotation 120 degrees, vertical travel approx.
20 inches), shall fature image rotation and reversal. Beta Mobile
Imaging Table model S 4015 shall be welded anodized aluminum frames
(top and bases), shall have a patient surfaces of approximately 26
inches X 75 inches (66cm X 190.5 cm), imaging surface shall measure
approximately 75 inches X 22 inches (190.5 cm X 55.8 cm), vertical
height shall range from 25 inches to 35 inches (10 inches), 63.5 cm to
88.9 cm (25.4 cm), the table shall allow access for the c-Arm opening,
the opening shall measure at least 48 inches X 37 inches (122 cm X 94
cm), shall be equipped with mounted oxygen bottle holder. The
Digimax-2000 High Definition Digital Imaging Processor shall feature
real time windows NT Graphic software with 12 Bit pipeline processor,
1024 X 512 X 8 bit image matrix, shall have 32 MB RAM expandable to 128
MB RAM, shall have 540 MB hard drive, which can be expandable up to 12
GB, the system shall feature Real-Time digital fluoror and digital
spot radiography, real-time subtraction and road mapping with real-time
window/level control shall be available, variable rate shall be up to
30 FPS for Cine Loop acquisition and playback, the system shall have 21
inch B/W or color monitor, shall feature split image screen display for
live and reviewing image comparison on one monitor, shall feature
measurement and quantitative analysis menu, with Alpha/Numeric keyboard
for menu selection and data entry, shall have patient information and
file management, the contractor shall provide maintenance (labor and
parts) and keep the equipment in good operation condition. NIH believes
that the above referenced firm is the only source that can provide this
requirement. The contract shall be for one year with two options for an
additional year. Any other firm that feels they are qualified to
provide such services must submit by the response date (15 calendar
days from date of this publication) a detailed capability statement
documenting their ability to provide this service by meeting the
requirements stated above. Information furnished shall include enough
material to perform a proper evaluation. Respondents will not be
notified of the evaluation results. Closing date will be 15 calendar
days after publication of this notice. See number Notes 26 Posted
10/26/98 (W-SN265371). (0299) Loren Data Corp. http://www.ld.com (SYN# 0232 19981028\66-0011.SOL)